CEM BX-500 Portable Infrared Calibrator, 122 to 932F

Representative image only.
CEM Mfr # BX-500 – Item # WZ-16111-00
Calibrate infrared thermometers on the go
  • Large 2-1/4" (57 mm) black body with 0.95 emissivity satisfies most test geometry requirements
  • Self-test upon startup alerts you to malfunctions, saving time
  • Dual-line LED shows set point and actual temperatures simultaneously

£965.00GBP / Each
£1,158.00GBP (incl VAT)

SPECS & DESCRIPTION

Specifications & Description

  • Max Temperature (° F)932
  • Min Temperature (° F)122
  • Temperature Resolution (° F)0.1
  • Temperature Accuracy±1.0⁰F at 212⁰F (±0.5⁰C at 100⁰C); ±3.6⁰F at 932⁰F (±1.8⁰C at 500⁰C)
  • Heat-up Rate40 minutes
  • Cooling Rate45 minutes from max to 100⁰C
  • Stability±0.2⁰F at 212⁰F (±0.1⁰C at 100⁰C); ±0.6⁰F at 932⁰F (±0.3⁰C at 500⁰C)
  • Width (cm)18
  • Height (cm)11.4
  • Depth (cm)23.3
  • Power (VAC)110 / 220
  • Power (Hz)50/60
  • DescriptionPortable Infrared Calibrator, 122 to 932F
Two year warranty

Includes

power cord.

Key Features

  • Large 2-1/4" (57 mm) black body with 0.95 emissivity satisfies most test geometry requirements
  • Self-test upon startup alerts you to malfunctions, saving time
  • Dual-line LED shows set point and actual temperatures simultaneously

More About this Item

This infrared calibrator features a microprocessor-based PID temperature controller with precision 100 Pt RTD providing accuracy and stability. Swing down the riser to raise the front end for greater visibility and easier programming. Switch between °F and °C to accommodate any thermometer or test protocol. Fan cools the heater quickly to a lower set point, so you can perform more calibrations.

REVIEWS

Q&A

Do you have questions about this product?

Get answers from real customers and in-house experts with AnswerBox.

1 Question | 1 Answer
Displaying question 1
  • Q:

    Hello,
    All of our other tools that we use to verify thermometers are sent out annually to be recalibrated by a third party vendor. Does this equipment require this also. If not, what is the justification.
    Thanks
    Asked on 6/7/2016 by Scott from Tyson Foods, Traverse City MI

    1 answer

    • STAFF EXPERT

      A:

      Just like all your other tools, it would need to be calibrated on a regular basis to make sure that it is working correctly.

      Answered on 6/7/2016 by Armen from Technical Support